ConformsTo
IEC61347-1 Annex D
Heating Tester of Thermally Protected Lamp Controlgear (Model:SFT S2-1215)
Application
Requirements for carrying out the heating tests of thermally protectedlamp controlgear.
ConformsTo
IEC61347-1 Annex D
Specification
A, Chamberinternal dimensions: 610*610*610 mm
B, Theentire test enclosure shall be constructed of heat resistant material 25 mmthick.
C, The floorof the test compartment shall measure 560 mm × 560 mm
D, Testplate material: insulation asbestos
E, The heat source used for the testenclosure described above shall consist of four 300 W strip heaters havingapproximate heating surface dimensions of 40 mm × 300 mm
F, Heating: start heating when thetemperature rises fast, power is big, when temperature close to set thetemperature of the heating power.
G,Temperature control mode: PID automatic temperature control.
H, Display:digital display
I, Range oftemperature: room temperature to 60 °C
J, Testtemperature: 40 °C +0/-5°C
K.Temperature uniformity: 2°C
L.Temperature volatility: + /-1 °C
M. Sample power supply: Electricalconnections may be brought out of the enclosure through the 150 mm squareopening illustrated in Figure D.1.
N.Thermocouple temperature measurement.
O. Outer Dimensions:W*D*H=700mm*650mm*800mm
P.Weight: 65KG
OrderingInformation
SFT S2-1215Heating Tester of Thermally Protected Lamp Controlgear